Basis: An air pump draws ambient air at a constant flow rate into
a specially shaped inlet where particulate matter is separated into
size fractions. Particulate matter is then collected on a filter. Each
filter is weighed before and after use, to determine the net mass gain
due to collected matter. The total volume of air filtered is known from
the constant air flow, and the difference in filter weights is used to
calculate the particulate matter concentration in micrograms per cubic
meter (µg/m3) of air.
Range: The concentration range is 30 to 300 µg/m3
for test conditions. (Ref. 24, Para. 53.34, Table C-4)
Lower detection limit, or lower quantifiable limit: This limit is
not set but is permitted to vary according to filter technology. (Ref. 25,
Para. 3.1)
Temperature tolerance: Samplers must operate over a -30° to
+45°C temperature range (Ref. 28, Para. 7.4.7). Some devices are
offered with temperature protection features, which are optional.
Known Interference:
Particulate matter may be lost during filter handling and weighing
procedures, especially if filter is exposed to warming.
Gaseous species may contaminate filters.
Humidity and absorbed water may be difficult to control both during
operations and when handling filters.
Removing filters and transporting to a lab for analysis may affect
results.
Meteorological conditions may affect flow rate.
Notes of Interest:
Precision (means agreement among successive measures of equal
samples) is defined by EPA as follows: must be 5 µg/m3
for PM10 concentrations below 80 µg/m3 and 7 percent for
PM10 concentrations above 80 µg/m3 (Ref. 25, Para. 4.1),
but the wording probably means "at least 5 µg/m3."
This requirement is simplified in Reference 26 (Para. 53.40, Table D-1)
to 5 µg/m3 or 7 percent for three collocated samplers.
Comparability: A procedure for testing comparability of PM10 methods
is provided. This procedure requires operation of the various devices
in the same context (immediate vicinity and time), and a subsequent
statistical analysis of measurements. The results are required to
correlate 0.97 with previously established reference, with respect to
sample masses measured over 24 hours (Ref. 24, Para. 53.34, Table C-4).
Upper capacity limit: All samplers should be capable of measuring
24-hour mass concentrations of at least 300 micrograms per standard
cubic meter of air (Ref. 25, Para. 3.1). Filter exchange features can
extend the upper capacity for some devices, which use various automatic
filter replacement mechanisms to overcome limitations of excess matter
collecting on one filter. Thus a given device might be pre-loaded with
several filters which will be used sequentially.
Accurate air flow measurement is crucial to the determination of
concentration of particulate matter, and sampling devices employ various
systems for flow determination. For example, a mass sensor sends a signal
to a microprocessor, which makes a determination and either increases or
decreases fan speed. Appendix J requires that flow rate be accurate to
within 2 percent of a preselected standard over a range of (potential)
ambient conditions (Ref. 25, Para. 8.2.2).
Basis: Beta particles (electrons with energies in the 0.01 to 0.1 MeV
range) are attenuated according to an approximate exponential function
when they pass through particulate deposits on a filter tape. Automated
samplers (analyzers) use a continuous filter tape, first measuring the
attenuation by the unexposed tape, and then measuring the attenuation
after the tape has passed through the ambient air flow. The attenuation
measurement converts to a measure of the mass on the filter, so that
the filters do not require later laboratory analysis for the mass
variable. For some devices, the beta particle source is 14C.
Range: The concentration range is 30 to 300 µg/m3
for test conditions. (Ref. 24, Para. 53.34, Table C-4)
Lower detection limit, or lower quantifiable limit: This limit is not
set but is permitted to vary according to filter technology. (Ref. 25,
Para. 3.1). An approximate value of 5 µg/m3 is given
in Ref. 27.
Temperature tolerance: Samplers must operate over a -30° to
+45°C temperature range (Ref. 28, Para. 7.4.7). Some devices are
offered with temperature protection features, which are optional.
Known Interference:
Particulate matter may be lost due to filter tape advance and
vibration, especially if filter is exposed to warming.
Gaseous species may contaminate filters.
Humidity and absorbed water may be difficult to control during
operations.
Meteorological conditions may affect flow rate.
Although on-site real-time mass measurement offers significant
improvements over the filter removal and laboratory analysis process,
the beta emission and detection process present additional on-site
maintenance requirements.
Notes of Interest:
Precision (means agreement among successive measures of equal
samples) is defined by EPA as follows: must be 5 µg/m3
for PM10 concentrations below 80 µg/m3 and 7 percent for
PM10 concentrations above 80 µg/m3 (Ref. 25, Para. 4.1),
but the wording probably means "at least 5 µg/m3."
This requirement is simplified in Reference 26 (Para. 53.40, Table D-1)
to 5 µg/m3 or 7 percent for three collocated samplers.
Comparability: A procedure for testing comparability of PM10 methods
is provided. This procedure requires operation of the various devices
in the same context (immediate vicinity and time), and a subsequent
statistical analysis of measurements. The results are required to
correlate 0.97 with previously established reference, with respect to
sample masses measured over 24 hours (Ref. 24, Para. 53.34, Table C-4).
Upper capacity limit: All samplers should be capable of measuring
24-hour mass concentrations of at least 300 micrograms per standard
cubic meter of air (Ref. 25, Para. 3.1).
Because of the characteristics of the Beta particles emitters,
which emit particles in irregular intervals, or bursts, the accuracy
of the devices improves with longer periods of sampling, during which
the emission frequency approaches a mean value, and thus detects the
"true" ambient concentration, as deposited on the filter. By this same
phenomenon, the detection of the "true" ambient concentration will be
detected earlier, and more accurately for higher concentrations.
Tapered Element Oscillating Microbalance (TEOM) Methods (FEM)
Basis: Air is drawn through a tapered glass element with a filter
attached. The element oscillates according to a characteristic frequency,
that decreases as mass accumulates on the attached filter. Measurement of
the change in frequency converts to measurement of the accumulated mass.
Range: The concentration range is 30 to 300 µg/m3
for test conditions. (Ref. 24, Para. 53.34, Table C-4).
Lower detection limit, or lower quantifiable limit: This limit is not
set but is permitted to vary according to filter technology. (Ref. 25,
Para. 3.1). An approximate value of 5 µgm3 is given in
Ref. 27.
Temperature tolerance: Samplers must operate over a -30° to
+45°C temperature range (Ref. 28, Para. 7.4.7). Some devices are
offered with temperature protection features, which are optional.
Known Interference:
Particulate matter may be lost due to vibration, especially if
exposed to warming.
Gaseous species may contaminate filters.
Humidity and absorbed water may be difficult to control during
operations.
Meteorological conditions may affect flow rate.
Although on-site real-time mass measurement offers significant
improvements over the filter removal and laboratory analysis process,
the TEOM equipment presents additional maintenance requirements.
Notes of Interest:
Precision (means agreement among successive measures of equal
samples) is defined by EPA as follows: must be 5 µg/m3
for PM10 concentrations below 80 µg/m3 and 7 percent for
PM10 concentrations above 80 µg/m3 (Ref. 25, Para. 4.1),
but the wording probably means "at least 5 µg/m3."
This requirement is simplified in Reference 26 (Para. 53.40, Table D-1)
to 5 µg/m3 or 7 percent for three collocated samplers.
Comparability: A procedure for testing comparability of PM10 methods
is provided. This procedure requires operation of the various devices
in the same context (immediate vicinity and time), and a subsequent
statistical analysis of measurements. The results are required to
correlate 0.97 with previously established reference, with respect to
sample masses measured over 24 hours (Ref. 24, Para. 53.34, Table C-4).
Upper capacity limit: All samplers should be capable of measuring
24-hour mass concentrations of at least 300 micrograms per standard
cubic meter of air (Ref. 25, Para. 3.1).
Basis: An air pump draws ambient air at a constant flow rate into
a specially shaped inlet. Suspended particulate matter is then collected
on a filter. Each filter is weighed before and after use, and moisture
content of filters is controlled. The total volume of air filtered is
known from the constant air flow, and the difference in filter weights
is used to calculate the particulate concentration micrograms per cubic
meter of air (µg/m3).
Range: The concentration range is 10 to 200 µg/m3
for test conditions. (Ref. 24, Para. 53.34, Table C-4)
Lower detection limit, or lower quantifiable limit: FRM samplers
should be able to detect a 24 hour average concentration of 2
µg/m3 (Ref. 28, Para. 3.1).
Temperature tolerance: Samplers must operate over a -30° to
+45°C temperature range (Ref. 28, Para. 7.4.7). Some devices are
offered with temperature protection features, which are optional.
Known Interference:
Particulates may be lost during filter handling and weighing
procedures especially if filter is exposed to warming.
Gaseous species may contaminate filters.
Humidity and absorbed water may be difficult to control both during
operations and when handling filters.
Removing filters and transporting to lab for analysis may affect
results.
Meteorological conditions may affect flow rate.
Notes of Interest:
Precision for PM2.5 samplers is specified as a 10 percent
coefficient of variation or better (Ref. 28, Para. 5). A separate method
of calculating precision for particulate samplers is given in Ref 26
(Para. 53.43(c)).
An upper limit for filter mass loading is not given. All samplers
are probably able to measure an ambient air concentration of at least
200 micrograms per cubic meter over a 24 hour period (Ref. 28, Para. 3.2).